HCP
page 28
40°C
34°C
ASF alarm
ASF activated
ASF activated
ASF aktiv
°C
ASF set to
+/-3°C in
+/-3°C in
SETUP
12.2.4 Automatic temperature monitor (ASF)
12.2.4 Automatic temperature monitor (ASF)
A monitoring device which automatically follows the selected temperature setpoint.
Automatic temperature monitor
OFF
(ASF OFF)
°C
MIN
AUTO
MAX
Using the push/turn control select
the symbol
AUTO
.
Hold down the SET key and select
off
using the push/turn control.
Automatic temperature monitor
ON
(ASF ON)
°C
MIN
AUTO
MAX
Using the push/turn control select
the symbol
AUTO
.
Hold down the SET key and select
on
using the push/turn control.
Notes on the ASF:
The tolerance margin for the ASF is selected in SETUP (see the menu item Tolerance margin ASF in the
section 10).
Tolerance margin reached = ASF activated
Tolerance margin reached = ASF activated
The ASF is automatically activated when the actual temperature has reached 50% of the selected tolerance
margin of the setpoint (in the example 37°C, -1,5°C).
The activation of the automatic temperature monitor is indicated by the bright symbol
AUTO
.