Chapter 9
– Errors
54
Error message
Possible causes
What you can do
エラー
43
:
カンソウ カアツ
ホ゜ンフ゜
Flow filter soiled
Check flow filter lower front on the chamber
(right behind the door)
On repeated occurrence notify your specialist
dealer
エラー
51
:
メッキン
TU2
Shortfall of minimum
permissible sterilization
temperature (temperature
sensor 2)
Operate device with smaller load, possibly
carry out Vacuum test.
Check door seal for wear.
On repeated occurrence notify your specialist
dealer.
Also see
Error 34
エラー
52
:
メッキン
TO2
Exceeding the maximum
permissible sterilization
temperature (temperature
sensor 2)
Carry out vacuum test
On repeated occurrence notify your specialist
dealer.
Also see
Error 35
エラー
53
:
メッキン
TD2
Maximum permitted difference
between theoretical
temperature, quoted from the
pressure signal, and the
temperature measured at the
temperature sensor 2 was
exceeded
On repeated occurrence notify your specialist
dealer.
Also see
Error 38
No display on the screen
After switching on the steam sterilizer the display remains empty.
What you can do
Check whether the power plug is correctly in the socket.
Check the supply voltage at the socket.
If necessary, switch the instrument securing at the lower front of the
steam sterilizer (see page 10,
Fig. 1/(7)
)off. Follow the instructions in
the technical manual in
Exchange device fuses
.
Too large feed water consumption
The feed water consumption is dependent upon the program and the
loading of the steam sterilizer.
What you can do
Check the correct set-up of the steam sterilizer.
The floor of the chamber must be free. Where applicable, remove
fallen instruments, filter paper, etc.