7-68
Code
Cl.
Operational
message
Description
Input
41
W
Setup is faulty
System of conditions on
master inhibits start of HV
operation if user level is
inappropriate.
Software trouble - must
not occur during opera-
tion
42
W
Access denied
Entered password is
wrong and will be
rejected.
To administrator,
customer service und
manufacturer: Enter
correct password!
43
W
Access released
Entered password is right
and will be accepted.
To administrator,
customer service und
manufacturer: Access to
extended Setup is
granted.
44
W
Not accepted
Entered password is
illegal and will be
rejected.
To administrator,
customer service und
manufacturer: Enter
approved password!
45
W
Relaxed mode
on
The slave is operating in
Relaxed mode after the
respective adjustment
has been made in Setup.
Must not occur during
operation - Relaxed
mode is reserved for
customer service!
46
W
DEMO MODE ON
The system has been set
to demo mode via the
setup menu.
Should not appear during
regular operation. Demo
mode is only for service
purposes.
47
W
C-values stored
Calibration data have
been changed by entry
and stored on slave in
non-volatile mode.
Any change of calibration
data is reserved for
customer service and
manufacturer !
48
W
Not calibrated
System has not been
calibrated / calibration
data in slave have been
lost (possibly only in
parts).
Perform initial calibration
/ Call customer service,
as initial calibration is
reserved for customer
service!
49
W
No print data
No data are availabe for
built-in protocol printing.
Record protocol, data will
be available till next
"HVOn".
50
W
Printer error
The built-in protocol
printer is not ready for
operation.
Check protocol printer
and its linkage as well as
power supply
51
W
IIC-Bus error
Some error has occurred
when reading from or
writing on the real-time
clock.
Access to the real-time
clock is via I²C Bus in
protocol node. System
card involved?