7.5.3 Operational Messages about Switching Conditions
The following operational messages inform the user about the switching conditions prior to
"HVOn" and about the cause of shut-down after "HVOff". These messages need to be
confirmed by the operator with "OK" unless they disappear on their own accord after
having set a switching condition.
Code
Cl.
Operational
message
Description
Input
-
C
Checking...
After the parameters have
been set, the switching
conditions are being
queried.
No action required,
only information
-
C
HV interlock
Detachable key switch is
switched off.
Switch detachable key
switch on
-
C
Emergency switch
Internal emergency switch
has tripped.
Switch emergency
switch on
-
C
Ext. safety loop
External safety loop is
open.
Close emergency loop
-
C
Door contact(s)
At least one door contact
is open.
Close doors of vehicle
-
C
Module coupling
Connection between
modules is interrupted..
Connect modules
-
C
SF6 pressure
SF6 gas pressure in HV
switching system is below
rated value.
Gas pressure monitor
at present not funct-
ional, check signal on
module
-
C
Humidity sensor
Degree of dew condens-
ation on system is above
permissible limit.
Dew monitor at present
not functional, check
signal on module, if
need be
-
C
HV connector
The HV cable is not
properly connected at the
HV connector
Connect the HV cable
-
C
Ground monitor
Earth monitoring is
reporting inadequate
protective earth
connection.
Check the electrical
connections /
grounding conditions
-
C
Cable shield
Earth monitoring is
reporting inadequate
operational earth
connection.
Check the electrical
connections /
grounding conditions