7-64
7.5.4 Error Messages and Warnings
With the exception of operating errors, every error, once it has occurred, is reported by
way of a two-digit error code. Each of these messages needs to be confirmed by the
operator with "OK", further procedures are determined by the system.
Code
Cl.
Operational
message
Description
Input
00
F
Unknown error
An error message has
been called that is
unknown to the system.
Software trouble -
must not occur during
operation
01
F
Event overflow
The software buffer of
events in the master has
overflown.
Software trouble - may
occur during operation
only if there is an unusual
accumulation of errors
02
F
Slave not found
The master has not
received any acknow-
ledgement of receipt from
the slave via LON.
LON Bus / wiring / power
supply
03
F
Slave not ready
The master has not
received any confirmation
from the slave on its
request.
LON Bus / wiring / power
supply
04
F
Unknown slave
The software version of
the slave is not com-
patible with the software
version of the master.
Insert Slave PROM with
suitable software version
05
F
Master timeout
Communication between
master and slave is inter-
rupted (slave does not
respond).
LON Bus / wiring / power
supply
06
F
P-node missing
A protocol node
registered in Setup does
not answer to the
network.
Check protocol node and
its linkage as well as
power supply
07
F
Config. restart
The system needs to be
rebooted after the device
configuration has been
changed.
Initiate restart by
confirmation of error
message
08
F
State mismatch
The slave is unable to
carry out the requested
transition of condition.
Software trouble - must
not occur during
operation