PERSONAL PROTECTION TESTS, 230 V SINGLE PHASE:
1. RCD 30 mA AC TEST
Test current:
31.5 mA AC (5% above nominal trip current)
Max. test time:
300 ms
Polarity selection:
0° and 180° of an input sine wave
2. RCD 6 mA DC TEST
Test current ramp
up at rate of 6 mA in 2.5 s, then held at 6 mA DC
Ramp polarity
positive and negative ramp
Max. test time:
12.5 s
Polarity selection:
0° and 180° of an input sine wave
PERSONAL PROTECTION TESTS, 120 V SINGLE PHASE / 240 V TWO PHASE
1. GFCI/CCID 5 mA AC test
Test current:
6 mA AC
Max. test time:
5.59 s
Polarity selection:
0° and 180° of an input sine wave
2. GFCI/CCID 20 mA AC test
Test current:
21 mA AC (5% above nominal trip current)
Max. test time:
5.59 s
Polarity selection:
0° and 180° of an input sine wave
NUISANCE TRIPPING TESTS, 230 V SINGLE PHASE
1. RCD AC ramp test
Test current
AC current ramp up to 30 mA in 2 mA steps
Step time:
300 ms
Max. test time:
4.5 s
2. RCD DC ramp test
Test current ramp
up at rate of 6 mA in 2.5 s, then held at 3 mA DC.
Max. test time:
11.25 s
NUISANCE TRIPPING TESTS, 120 V SINGLE PHASE / 240 V TWO PHASE
1. GFCI/CCID 5 mA AC test
Test current
AC current ramp up to 6 mA in 0.5 mA steps
Step time:
100 ms
Max. test time:
1.2 s
2. GFCI/CCID 20 mA AC test
Test current
AC current ramp up to 20 mA in 1 mA steps
Step time:
100 ms
Max. test time:
2 s
TOUCH VOLTAGE TEST
Options:
25 V or 50 V limit
Test current typically 1/3 of RCD nominal test current