2.
Product Specifications
2.1.
Modes
-
Automatic shutoff of the pressure monitor if no CPAP use is taking place
-
21–100% continuous adjustment of the oxygen concentration
-
Reset for alarms – 2-minute alarm silencing (make silent or end):
Warning – Alarm Silencing
Silencing the alarm requires additional monitoring of vital signs!
-
Confirmation of the nCPAP alarm setting for automatic monitoring
-
Confirmation of the FiO
2
concentration setting for automatic monitoring (± 10% of the O
2
)
-
Disconnection or leakage alarm (if the patient pressure falls below 1.5 cm H
2
O) - Alarm after 3
seconds
-
Battery monitoring with charge status display and alarm
-
Overpressure alarm in 3 seconds (if the patient pressure exceeds 10 cm H
2
O)
-
Microprocessor-controlled input of programmes on FLASH (= technical upgrade)
2.2.
Controls and Displays
-
Flow adjustment control knob
-
FiO
2
adjustment control knob
-
Soft-touch membrane keypad
-
Monitor with integrated 96 x 61 mm LCD display and illumination resolution 240 x 128 dots b/w
2.3.
Monitoring
-
Oxygen supply pressure, air supply pressure
-
Spontaneous respiratory rate
-
Alarm functions:
CPAP
FiO
2
Battery charge indicator
Mute setting
Automatic alarm system for loss of oxygen or air pressure:
acoustic and pneumatic
2.4.
Pneumatic Properties
-
Patient gas outlet: Connection with 22 M and 15 F converter
-
Patient flow: 0 – 16 L/min, working area 4-10 l/min
-
Gas supply: Oxygen and air
-
Range: 3.5 – 6.0 bar
-
Adjustment range: 21% to 100% oxygen (tolerance ±3%)
-
Gas connections: for DISS supply (alternative to NIST)
-
Oxygen sensor for FiO
2
measurement - sensor type MLF16 (unleaded)
-
Overpressure valve 80 mbar