9.2
Electromagnetic immunity
Guidance and manufacturer’s declaration - electromagnetic immunity
The medinSINDI
®
is intended for use in the electromagnetic environment specified below. The customer or user of
the medinSINDI
®
, or the organisation responsible, should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV electrostatic
discharge
± 8 kV air discharge
± 6 kV electrostatic
discharge
± 8 kV air discharge
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surges
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% U
T
for 1/2 cycle
(> 95% dip)
40% U
T
for 5 cycles (60%
dip)
70% U
T
for 25 cycles (30%
dip)
< 5% U
T
for 5 s (> 95%
dip)
< 5% U
T
for 1/2 cycle
(> 95% dip)
40% U
T
for 5 cycles
(60% dip)
70% U
T
for 25 cycles
(30% dip)
< 5% U
T
for 5 s (> 95%
dip)
Mains power quality should be that of
a typical commercial or hospital
environment.
Thanks to its built-in battery, the
medinSINDI
®
continues to operate
during power mains interruptions. It
therefore does not need to be
powered from an uninterruptible
power supply or external battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
The strength of power-frequency
magnetic fields should correspond to
that of a typical commercial or
hospital environment.
Note:
U
T
is the a.c. supply voltage prior to application of the test level