Attach device to wall/rolling cart rail mounting and secure in place:
-
Loosen the knurled head screws of the rail clamps on the back of the device
-
Hang device on wall rail or rolling cart rail
-
Tighten the knurled head screws.
Figure 1: Screws for attachment to the rolling cart rail
The medinSINDI
®
can be transported when it is not in use; however, when it is in use, it must be securely
mounted. However, if the medinSINDI
®
becomes damaged during this transport or due to rough handling, its
function must be checked by a service technician. It is not allowed to be used until then.
3.2.
Connection to Gas Supply
medinSINDI
®
requires clean, oil-free and dry air as well as medical oxygen; both gases need to be at an
operating pressure of 3.5 to 6.0 bar respectively.
The central gas supply or gas bottles with a connected pressure-regulating unit can be used as sources of air
and oxygen.
Air and oxygen must have the same pressure stage (maximum difference ± 1 bar). medinSINDI
®
monitors the
pressure of both gases to ensure this correct gas mixture.
medinSINDI
®
requires both gases simultaneously for the connection, otherwise an acoustic alarm will
immediately sound. (Loss of gas)
Important Note – Gas Supply
-
The oxygen content of the gas mixture can no longer be controlled if the difference in the supply
gas pressures is greater than 1 bar. In this case, an internal valve opens and connects the air and
oxygen supply channels to compensate for the lower gas supply. As of this point, the mixer loses
its function and an acoustic-pneumatic alarm is triggered.
-
Connect medinSINDI
®
properly to the central gas supply or full gas bottles (oxygen and air). This
will ensure high gas flows.
-
When using gas bottles, set the initial pressure regulator to a pressure between 3.5 and 5.0 bar
and check the alarm function.
-
Ensure that the pressure reducers of the gas supply tubes – if available (see list of accessories,
page 29) – are set to 3.5 bar.
-
Supply gases only via the connections provided.
3.3.
Gas Supply Connection (see Figure 3)
-
Screw air gas supply tube (b/w) onto medinSINDI
®
-
Screw oxygen gas supply tube (white) onto medinSINDI
®
-
Attach air gas supply tube (b/w) to central gas supply or full gas cylinder (3.5 – 6 bar)
-
Attach oxygen gas supply tube (white) to central gas supply or full gas cylinder (3.5 – 6 bar)
Knurled head screws