After setting the desired oxygen concentration with the control knob on the mixer, the alarm can be activated
to monitor the concentration by pressing the "CONFIRM FiO
2
" button. The alarm sounds in the case of
deviations of ± 10%.
Important Note – O2 Alarm
Any change in the alarm limits of the oxygen concentration must be made by pressing the “CONFIRM FiO2"
button, otherwise an alarm will be triggered. (± 10% of the concentration)
4.3.5.
Spontaneous Respiratory Rate
A sensitive nCPAP pressure measurement is necessary to determine the respiratory rate. The pressure
difference must be approx. ± 1 cmH
2
O during inhalation and exhalation.
The rate is determined, calculated and displayed in each case for 30-second intervals.
If no spontaneous respiration is detected, 2 lines are seen instead of the spontaneous respiratory rate display.
4.3.6.
Oxygen Concentration Adjustment Trend
Figure 6: Oxygen concentration trend
Starting with software version 1.019, the adjusted oxygen concentration of the past 8 hours can be displayed.
To do this, hold down the "CONFIRM FiO
2
" button for 5 seconds. After 20 seconds, medinSINDI
®
automatically
switches back to the regular CPAP monitor. If medinSINDI
®
is switched off, the saved FiO
2
values are deleted.
Starting with software version 1.023, trending is automatically discontinued in the event of an alarm.