+49 8142 448 460
[email protected]
www.medingmbh.com
14 von 36
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinSINDI EN
Rev 14 Stand 30.10.2017
3.5.
Connection to Medijet
Figure 3: Connection to Medijet
®
A.
Screw air gas supply tube (b/w) and oxygen gas supply tube (white) onto medinSINDI
®
.
B.
Connect air gas supply tube (b/w) and oxygen gas supply tube (white) to the central gas supply or a full
gas cylinder.
C.
Insert external power adapter plug into connector provided and screw in place.
D.
Connect short tube from the patient tube set on the outlet converter (REF 51091) and connect to
humidifier chamber - any type of humidifier chamber is suitable.
E.
Using the long tube from the patient tube set, connect the humidifier outlet to the inlet of the Medijet
®
nCPAP generator. Connect temperature probe and power cable.
F.
Connect the nCPAP pressure line on the Medijet
and guide back to the medinSINDI
®
. It will be
connected there at the "PATIENT" connection in order to measure nasal pressure and display on the
screen. The nCPAP connection to medinSINDI
®
is a female Luer-type. The pressure line can be found in
each patient tube set.
G.
Optimal manual push ventilation (SMPV)
3.6.
Connection to nasal cannula
The system is set up as described in chapter 3.5, A to D. The difference is in step E: the nasal cannulas are
connected instead of the Medijet
®
. Steps F and G are omitted.
Important note when using nasal cannula
-
During device start-up, nothing should be connected to the medinSINDI
®®
and also no flow should
be emitted.
-
After the device start, activate "alarm silent". Then "CPAP alarm off" appears in the upper right-
hand part of the display.
-
Now the therapy can be started and the flow as well as the concentration can be adjusted
accordingly.
E
A
C
D
F
F
G