10.
Appendix – Device Passport
Gerätepass/Device Passport
medinSINDI
®
REF 1080
SN:
CPU-Monitor SN:
Akku-SN:
Software Version:
Oxygen cell SN:
Blender SN:
Gas connection system:
DISS
NIST
Front panel:
sindi
Test equipment:
Oxygen
measurement
(IF)
Oxi Quant MC
(Envitec)
Flow analyser PF-302
LOW (IMT)
Flow analyser PF-300
(IMT)
Mass flowmeter
Model 4143 (TSI)
Manometer 0-25 cm WS
(Suchy)
Test of oxygen calibration:
21% ±3%
ok /
nok
40% ±3%
ok /
nok
60% ± 3%
ok /
nok
100% ±3%
ok /
nok
Test of pressure calibration:
0 cmH2O ± 0,5
ok /
nok
10cm H2O ± 0,5
ok /
nok
Overpressure valve:
ok /
nok
Alarm test:
Dissconnection
alarm (P<1,5cmH20
~ 5sec)
ok /
nok
High pressure alarm
P> 10cmH2O ~5sec
ok /
nok
nCPAP low alarm
(P< CPAP low alarm
~ 20sec)
ok /
nok
nCPAP high alarm
(P>CPAP high alarm
~5sec)
ok /
nok
Audio alarm
(dis)able
ok /
nok
Audio alarm off / on
ok /
nok
Oxygen high low /
alarm
ok /
nok
Gas supply pressure
Air/O
2
ok /
nok