Do not clean with chemical cleaning agents.
Maintenance/Repairs/Modifications:
The power supply unit FW7555M/24 of the medinSINDI
®
is maintenance-free. It should not be opened.
The power supply unit may only be repaired by authorized specialists.
Warning
To avoid an electrical shock, the housing of the power supply unit of the medinSINDI
®
should not be opened.
Modification of the power supply unit is not permitted. (Termination of the warranty)
6.6.
Maintenance
6.6.1.
Every 2 months - oxygen calibration:
Before the initial start-up and at least every 2 months, an oxygen calibration must be performed.
In the event of frequent or longer-term operation of medinSINDI
®
, oxygen calibration must be performed if
these criteria are met:
Up to software version 1.019 – if the control knob for setting the oxygen concentration is turned completely to
the
left
(21%)
and
an
oxygen
concentration
of
≤
19%
is
shown
on
the
display.
Starting with software version 1.020 – if the control knob for setting the oxygen concentration is turned
completely to the right (100%) and an oxygen concentration of ≤ 98% is shown on the display.
Important Note – Test for Oxygen Calibration
Before each use of medinSINDI
®
on a new patient, it is recommended that a test be conducted to determine if
oxygen calibration must be performed.
Important Note – Test for Oxygen Calibration
Never test while medinSINDI
®
is connected to a patient, since the oxygen concentration is changed in the
driving flow during this time.
6.6.2.
Maintenance intervals:
The manufacturer recommends preventive maintenance every 12 months (see maintenance interval sticker on
the right-hand side of the device housing). The oxygen cell (REF OOM102) is to be replaced and the electronics
are to be calibrated. In the case of the oxygen cell (REF MLF16) is a replacement needed every 3 years. The
internal rechargeable battery is to be replaced every 3 years. In case of doubt, please contact the manufacturer
or the sales and service partner authorised for your country.
Important Note – Maintenance
-
The maintenance of the device may only be performed by trained personnel authorised to
perform such maintenance!
-
The device must be disconnected from the patient
-
Check the mixing accuracy and the overall function of the device
-
Check the tube connections and the gas connector prior to each use
Note – Device Passport
For maintenance and after each opening, the device passport must be filled out and a new safety seal must be
affixed. (see appendix)