In the next step, the desired setting is selected using the two CPAP low-alarm buttons and confirmed using the
Confirm CPAP button.
Meaning of the Setting:
Acoustic Alarm off possible: In this setting, the acoustic alarm can be switched off during normal operation. If
the Alarm Reset button is only briefly pressed during normal operation, the acoustic alarm is silenced for 120
seconds and then automatically becomes active again. However, if the Alarm Reset button is held down for 5
seconds, a brief tone is heard and from this point, all alarms are only indicated visually and no longer
acoustically. This setting does NOT automatically reset itself; instead it must be reactivated by holding down
the Alarm Reset button for 5 seconds.
Acoustic Alarm off not possible: In this setting, the option of deactivating the acoustic alarm for a longer
period of time is not available. In normal operation, there is only the option here of silencing acoustic alarms
for 120 seconds by holding down the Alarm Reset button. Then the alarms are automatically acoustically
reactivated. Holding down the Alarm Reset button for 5 seconds does not have any effect on the device.
4.4.3.4. Internal Data
Figure 13: Internal Data
The current software version can be found in this file. This is used to identify the CPU system and the special
functions for the measuring transducer and the O
2
cell. The CPU works with "Flash Memory". At the customer's
request, medin can update the programme with the most recent software version at any time in the Puchheim
plant.
To close this menu, hold down the “Confirm FiO
2
” button for more than 2 seconds.
4.4.3.5. Restarting after Calibration
The service menu can be left by selecting the "Exit" item. All calibration data are stored in the internal memory.
medinSINDI
®
now starts automatically or can be started using the "ON" button, at which time the system check
is performed.
4.4.4. External Alarm
Starting with software version 1.016 it is possible to combine medinSINDI
®
with an external alarm feature. For
this purpose, the conversion kit for medinSINDI
®
central control/ nurse alarm (accessory REF 11810) must
either be retrofitted or medinSINDI
®
must be ordered with this modification already included. This external
alarm feature allows alarms from medinSINDI
®
to be indicated via a central ward alarm.
The cable supplied by medin must be used to connect the medinSINDI
®
to an external alarm system in order to
achieve EMC according to DIN EN 60601-1-2:2007 (REF 11810) and the connection must be set up initially by an
appropriately trained technician.
Attention
The medinSINDI
®
may only be connected to an external alarm system using the connection cable provided by
the manufacturer (REF 11810), otherwise the functionality and EMC of the device cannot be guaranteed.