Inside the upper cover
Service Manual - medifa 8000
8
After removing the expansion bellows from the upper cover,
do the same with the control panel.
Do as follows: Loosen the screws (4 screws) from the inside
of the upper cover.
Now move the control panel towards the headboard, pull the
screw heads through the notch and remove the control panel.
fig. 20: removing the control panel
fig. 21: removing the data cable of the
control panel
Put the visible data cable off the jack.
The control panel is now dismantled and can be put aside.
Now the expansion bellows can be pulled over the frame. A
fig. 19: removing the expansion bellows
fig. 22+23: remove data cable from the
control panel
Now, remove the cables from the illuminated push button,
from the inside.
Just pull each plug off the illuminated push button.
Note:
Keep in mind the plug-in location and order as well as the
respective colours, to prevent a mix up.