Finally, sepeaate the wire of the way measurement system
( Trendelenburg shifting) .
Remove the two screws and open the way measurement
cap. You will now see the plug. Simply put it off.
After separating the way measurement system from the cylin-
der and cutting off the connection, simply screw off the holder
of the driving pin.
Remove the two fastening screws and mount the holder di-
rectly to the driving pin of the new system.
Note
:
Should the cable be faulty, too, retrace it back to the connec-
tion and replace it.
See the diagram for the correct placement of the cables in
the breaker box.
20
fig. 68: looseningholder of driving pin
fig. 66+67: