XV. Wiring Diagram Breaker Box (black)
Service Manual - medifa 8000
X1—way measurement trendelenburg (Wegmesser Trendelenburg)
X1.1 +5V DC
(gn)
X1.2 signal
(br)
X1.3 GND
(ws)
X1.4 PE
(rt)
X2—way measurement lateral (Wegmesser Lateral)
X2.1 +5V DC
(gn)
X2.2 signal
(br)
X2.3 GND
(ws)
X2.4 PE
(rt)
X3—way measurement transversal shift (Wegmesser Querverschiebbarkeit)
X3.1 +5V DC
(gn)
X3.2 signal
(br)
X3.3 GND
(ws)
X3.4 PE
(rt)
X4—way measurement longitidunal shift (Wegmesser Längsverschiebbarkeit)
X4.1 +5V DC
(gn)
X4.2 signal
(br)
X4.3 GND
(ws)
X4.4 PE
(rt)
X5— way measurement height adjustmend (Wegmesser Höhenverstellung)
X5.1 +5V DC
(gn)
X5.2 signal in
(br)
X5.3 GND
(ws)
110 ohms resistor between signal in and GND must be removed (110 Ohm-Widerstand zwischen GND und Signaleingang
entfernen) (rt)
X6—joystick (Joystick)
X6.1 +5V DC
(ws)
X6.2 joystick y-axis (Y-Achse Joystick)
(gn)
X6.3 joystick x-axis (X-Achse Joystick)
(ge)
X6.4 GND
(br)
X6.5 Schirmung Joystickkabel
X7—X10 to sub-d connectors (zu den Sub-D-Einbausteckern/Buchsen)
X7 control box (Steuerung)
X8 handswitch (Handschalter)
X9 column keyboard (Säulentastatur)
X10 diagnostic (Diagnose)
X12 Erdungsanschluss
X11
X