Service Manual - medifa 8000
Before being able to loose the expansion bellows, it is necessary to dismantle the crusting protector of the sli-
ding function.
Please see
Chapter IV „Removing the Crusting Protective Cover“
.
Do as follows:
Stand in front of the table.
Loosen the front holders of the expansion bellows by remo-
ving the two hexagon bolts on the upper side of the lower
base plate.
Now move to the back of table and remove the two other he-
xagon bolts on the upper side of the lower base plate.
After loosening all fastening screws, the expansion bellows
and holders can be left on the remaining holders on the co-
lumn.
The table construction is now accessible.
VI. Loosening The Expansion Bellows
10
fig. 30: loosening the holders of the ex-
pansion bellows
fig. 31 : loosening the holders of the
expansion bellows
fig. 32: : loosening the holders of the
expansion bellows