After removing the table top, you can start to remove the in-
tegrated crusting protector,
Move the table top with the help of the cross-sliding function-
to the complete right, ( from the patients side of view).
You will find the connection for the crusting protector of the
sliding mechanism.
By loosening the screws you can pull the cables from the
socket.
The cables can now be pulled off easily from the cable ducts
on the surface of the base plate.
If necessary, move the runners for the longitudinal shifting.
.
Attention:
Do not pull the slides out of the guide rail.This will de-
stroy the guide rail and will make it useless!
Therefore we recommend to protect the guide rail with a
cable tie against the complete pull out. ( Pull the cable tie
through the hole in the middle of the guide rail, while it is
in the middle position!)
Now loosen the cover panel of the crusting protector. Remo-
ve all screws on the front and on the back.
Now the cover panel can be bend and removed.
Attention:
Before removing the last screw, a second person should
secure the cover panel to prevent it from falling down!
Now you will see a second cover panel, as well kept together
with two screws, but, in addition, secured by eight more
screws, screwed from below.
Loosen all screws and remove the cover panel by bending it
up!
Attention:
Before removing the last screw, a second person should
secure the cover panel to prevent it from falling down!
IV. Removing The Crusting Protective Cover
medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77
5
fig. 7: remove crusting protector
fig. 6: remove crusting protector
fig. 10: remove crusting protector
fig. 8+9: remove crusting protector