Before replacing the way measurement system for the longitudinal shifting the table top has to be taken off.
For further help please see
Chapter III. „Removing the Table Top“.
Do as follows:
After taking off the table top, you will be able to look at the
way measurement system and dismantle it.
To replace the way measurement system just remove the
screw at the end.
Now retrace the cable to the connection of the expansion
bellows holder.
There, the cables can be removed by opening the cable
clamps.
X. Replacing the Way Measurement System
(Longitudinal Shifting)
medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77
17
fig. 58: dismantle way measurement sys-
tem
fig. 60: dismantle way measurement sys-
tem
fig. 59 dismantle way measurement sys-
tem