![Medifa 8000 Series Service Manual Download Page 16](http://html.mh-extra.com/html/medifa/8000-series/8000-series_service-manual_1761124016.webp)
Service Manual - medifa 8000
Do as follows:
First, screw back the cylinder connection to the lower base
plate.
Properly place the base plate onto the guide rails. Now move
to foot position of the table.
Take care that the position screw, which you will see on the
lower base plate,exactly fits into the connection (signal gene-
rator) of the way measurement system for the transverse shif-
ting.
Improper handling might lead to serious problems of the shif-
ting functions.
As soon as the upper base plate fits properly, the plate can
be fixed to the guide rails by screws.
In order to reach all screws move the plate by hand.
Attention:
Do not pull out the base plate completely out of the gui-
de rails. That would destroy the guide rail system and
render it useless!
Finally, fix the other cylinder connection to the back of the
upper base plate.
IX. Mounting of the Upper base plate
16
fig. 56+57:
fig. 53: fastening hydraulic cylinder
connection
fig. 54+55: signal generator ( way measu-
rement )