Changing Rate During Delivery
You can change the infusion rate during any delivery session, except Volume/Time or Intermittent modes.
1.
To change the rate during delivery, press the
Chg Rate
(or
Chg Dose
) button.
2.
Use the number buttons to set the dose rate. Rate is limited by the maximum and minimum delivery rate for
syringe size.
3.
When ready press the
Enter
button to accept the new setting. The pump delivery rate does not change until
you press the
Enter
button. If you change your mind, press Back to return to D
ELIVERY
screen without
making a change.
Changing Rate When Paused
You can change the dosage rate at any time the pump is paused:
1.
If the pump isn’t already paused, then press the
Stop
button to pause the infusion.
2.
To change the rate during any pause, press the Chg Rate (or Chg Dose) button.
3.
Use the number buttons to change the delivery rate. Rate is limited by the maximum and minimum delivery
rate for syringe size.
4.
When ready press the
Enter
button to accept the new setting.
a)
The delivery rate entry does not change until you press the
Enter
button.
b)
You may exit without changing the rate by pressing the
Back
button.
5.
The pump is paused at the B
EGIN
I
NFUSION
screen. When ready, press the
Start
button to begin delivery.