M
EDEX
Medfusion 3500
Medex Operation Manual
Artwork GD000608 Revision 5
Page 37
C
ANCEL
– stops the bolus and returns to the normal delivery rate and screen.
S
TOP
– during a bolus delivery immediately halts all delivery and the pump pauses.
4.
The bolus delivery continues at the bolus delivery rate until the bolus volume is delivered. On completion
of bolus dosing … the pump beeps, then resumes the programmed infusion delivery.
5.
The bolus volume delivered is automatically added to the Total Volume Delivered (TVD) as well as the
Programmed Volume Delivered (PVD), or Programmed Dose Delivered (PDD) for that infusion.