Pump setup and syringe loading is a multi-step process, based on your daily practice of infusing patients with
medication or other fluids.
Turning on the Pump
Below are an illustration of the startup screen, and a list of basic steps for turning on the infusion pump.
1.
Press and hold the
Power
button until the dark screen appears. Verify the screen shows all pixels on, then
all pixels off.
2.
The pump should “beep” twice and then the indicators flash for Alarm, Infusing, and Lock as the self-test
cycles. The pump ID, if it has one, appears in the upper left corner. In this case, it is called “
Jobara
”. The
pump serial number will appear at the middle right, if it has been set in the software. In this case, it is
“M01234”.
3.
You should now verify the display indicates successful completion of system startup self-tests.
W
HAT IF THE
P
UMP
D
OES
N
OT
T
URN
O
N
?
If the pump does not turn ON with successful completion of self-tests, remove it from service and have it checked by
a trained biomedical
technician.