M
EDEX
Medfusion 3500
Medex Operation Manual
Artwork GD000608 Revision 5
Page 29
•
Dose Volume
•
Delivery Time
The pump calculates the correct flow rate by dividing the volume by the delivery time. Volume limit, loading or
bolus dose can
not
be set in this mode. Find and select V
OLUME
/
T
IME
from the S
ELECT THE
M
ODE
screen. You may
have to press the
More
button to find it.
1.
When ready, press the
Enter
button to accept the syringe. This calls the V
OLUME TO
I
NFUSE
Screen.
2.
At the V
OLUME TO
I
NFUSE
screen, press the
number
buttons to program the volume setting. Volume is
limited to the maximum fill volume for the syringe size – and minimum volume is 1/600th of the selected
syringe size.
3.
When ready, press the
Enter
button to go to the D
ELIVERY
T
IME
screen.
4.
At the D
ELIVERY
T
IME
screen press the
number
buttons to set delivery time. Time is always limited by
maximum and minimum rates for selected syringe size.
5.
When ready, press the
Enter
button to accept the setting to go to the B
EGIN
I
NFUSION
screen.
6.
All settings are made, and the pump pauses at the B
EGIN
I
NFUSION
screen. Confirm all settings are correct.
You may now accept the settings and start the infusion, or use the three menu buttons to modify the settings.
7.
Prior to beginning the infusion, press <BOLUS> to initiate the priming sequence (page 21).
8.
When ready, press the
Start
button to begin the infusion.