47
CLEANING AND STERILISATION
Rinse the inner channel of the scaler
handpiece using a 20 ml syringe
previously filled with a new enzymatic
solution. Repeat twice;
4
x2
Remove the handpiece, nose cone, and
light pipe from the enzymatic solution
and gently brush their surfaces using a
brush with soft nylon bristles under hot
(40 °C
±
5 °C) running water, for at least
10 minutes;
5
Fill the ultrasonic tank with the enzymatic
detergent solution prepared according to
the manufacturer’s instructions.
Use water at ambient temperature (22 °C
±2 °C);
6
Place the scaler handpiece, nose cone,
and light pipe in the ultrasonic tank
immersed in the enzymatic detergent
solution at 22 °C ±2 °C and run a cycle
for at least 10 minutes;
7
Summary of Contents for MICROPIEZO
Page 2: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figura 1 Componenti Components Composants...
Page 3: ...MICROPIEZO MANUALE D USO E MANUTENZIONE...
Page 4: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figura 1 Componenti MICROPIEZO...
Page 8: ...IV MICROPIEZO PAGINA LASCIATA INTENZIONALMENTE BIANCA...
Page 87: ......
Page 88: ...02150403 MICROPIEZO V IT EN FR Rev 07 del 31 05 2021...
Page 89: ...MICROPIEZO USE AND MAINTENANCE MANUAL...
Page 90: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figure 1 Components MICROPIEZO...
Page 94: ...IV MICROPIEZO PAGE INTENTIONALLY LEFT BLANK...
Page 173: ......
Page 174: ...02150403 MICROPIEZO V IT EN FR Rev 07 del 31 05 2021...
Page 175: ...MICROPIEZO MODE D EMPLOI ET D ENTRETIEN...
Page 176: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figure 1 Composants MICROPIEZO...
Page 180: ...IV MICROPIEZO PAGE INTENTIONNELLEMENT LAISS E EN BLANC...
Page 259: ......
Page 260: ...02150403 MICROPIEZO V IT EN FR R v 07 du 31 05 2021...
Page 261: ......