www.masimo.com
43
Masimo
Chapter 7: Specifications
Performance Specifications
Functional Oxygen Saturation (SpO2)
Condition
Range
A
rms
*
No Motion [1]
60% to 80% 3%
No Motion [2]
70% to 100% 2%
Motion [3]
70% to 100% 3%
Low Perfusion [4] 70% to 100% 2%
Pulse Rate (PR) [5]
Condition
Range
A
rms
*
No Motion
25 - 240 bpm 3 bpm
Motion [3]
25 - 240 bpm 5 bpm
Low Perfusion [4 ] 25 - 240 bpm 3 bpm
Total Hemoglobin (SpHb) [6]
Condition
Range
A
rms
*
No Motion 8 g/dL - 17 g/dL 1 g/dL
*The
A
rms
Accuracy is calculated based upon measurement values that are statistically
distributed; approximately 68% of the measured values fell /- the
A
rms
value when