Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Power frequency
(50/ 60 Hz)
magnetic field
IEC 61000-4-3
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of typical
location in a typical hospital
environment.
Portable and mobile RF
communications equipment should be
used no closer to any part of the ME
Equipment, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Conducted RF IEC
61000-4-6
3 Vrms 150
kHz to 80 MHz 3 Vrms
150 kHz to 80 MHz
Radiated RF IEC
61000-4-3
3 V/m 150 kHz
to 80MHz
3 V/m
80 MHz to 800 MHz
ISO 80601-2-61,
Clause 202
20 V/m 80
MHz to 2.5
GHz
20 V/m
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
, should be
less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol: