Pronto
Chapter 5: Messages
www.masimo.com
36
Masimo
Message
Explanation
Next Steps
INC DET
(Blinking)
Interference detected
Ensure that the sensor is properly
applied, and cover the sensor site with
opaque material, if required.
INC SEN
Incompatible sensor
Attach appropriate sensor.
INC CBL
Incompatible cable
Attach appropriate cable.
NO CBL
No cable
Attach appropriate cable.
RPL CBL
Cable life expired
Replace cable.
Temporarily blinking
message:
RPL
CBL
Cable life expired
Replace cable as soon as possible.
Temporarily blinking
message:
RPL
SEN
Sensor life expired
Replace sensor as soon as possible.
Temporarily blinking
message:
RPL
ADH
Adhesive sensor life
expired
Replace adhesive sensor as soon as
possible.
CHC SEN
Check sensor connection
Reattach sensor.
SEN 000
Zero sensor uses remaining
Attach new sensor. Dispose of the old
sensor per local governing ordinances.
See
Compliance Warnings and Cautions
on page 14.
RE TST
Spot check incomplete
Confirm sensor placement and press
SpHb
button again.