Chapter 8: Service and Maintenance
The following chapter contains information about cleaning, battery operation, performance
verification, service, repair, and warranty.
Cleaning
Pronto is a non-sterile and reusable device. The surface of the Pronto should be cleaned when
the device is visibly dirty, before and after each procedure, and/or according to hospital
practice.
To surface clean, wipe down the outer surface of Pronto using any of the following:
•
A soft cloth dampened with a mild detergent and warm water solution
•
Cidex Plus (3.4% glutaraldehyde)
•
10% bleach solution
•
70% isopropyl alcohol solution
Note
: Do not allow liquids to enter the interior of the device.
Note
: The performance of a device with a touchscreen will not be affected when using the
recommended cleaning solutions.
Battery Replacement
Pronto is powered by four (4) AA alkaline batteries. Do not use any other type of batteries or
power source to run the device.
Replacing the batteries
1.
Locate the battery compartment on the back of the device.
2.
Remove the battery cover by depressing the small rectangular button at the
bottom of the cover, and sliding the cover down off the bottom of the device.
3.
Remove the batteries and install new batteries in the directions indicated by the
battery orientation icons (+ and -) inside the battery compartment.
4.
Replace the battery cover by sliding it back up from the bottom of the device until
the rectangular locking button snaps back into position.
For more information about battery disposal, see
Compliance Warnings and Cautions
on page
14.