Pronto
Chapter 8: Service and Maintenance
www.masimo.com
58
Masimo
Performance Verification
To test the performance of the Pronto following repairs or during routine maintenance, follow
the procedure outlined in this section. If the Pronto fails any of the described tests,
discontinue its use and correct the problem before returning the device back to the user.
Before performing the following tests, verify or install new batteries into the Pronto. See
Battery Replacement
on page 57. Also disconnect any patient cables, serial cables, or sensors
from the device.
Power-On Self-Test
1.
Turn on the device by pressing the
Power
button. For about 5 seconds, all
available LEDs are illuminated and a brief beep tone sounds.
2.
Verify that the sensor uses remaining displays.
3.
Pronto is ready for use (the
RDY
message displays).
Service and Repair
Repair Policy
Masimo or an authorized service department must perform warranty repair and service. Do
not use malfunctioning equipment. Have the device repaired.
Clean contaminated and/or dirty equipment before returning, following the cleaning
procedure described in
Cleaning
on page 57. Make sure the equipment is fully dry before
packing.
To return the device for service, refer to
Return Procedure
on page 59.