Chapter 5: Messages
Messages
The following messages are specific to Pronto:
Message
Explanation
Next Steps
NO SEN
No Sensor Connected
1.
Connect sensor to cable.
2.
Check sensor connection to
cable.
3.
The Masimo rainbow
ReSposable® sensor
system is not connected to
the device
Assemble the ReSposable sensor system
and then connect to the device.
SEN OFF
Sensor off patient
1.
Reattach sensor to patient.
2.
Verify proper sensor
placement.
Circulating LEDs
Sensor is initializing/
determining measurement
Wait for pulse detection. (This search
should occur whenever a spot check is
performed). If necessary, shield the
sensor from excessive ambient or
strobing light.
Low SIQ Indicator
illuminates
Low Signal IQ
1.
Rule out occlusion of blood
flow.
2.
Verify placement of sensor.
3.
Move sensor to a better
perfused site.
4.
See
Low Signal I.Q. (Low SIQ)
on page 40.
Single Battery Level
Indicator flashes
(with audible alert)
Battery level too low
Replace batteries immediately. See
Battery Replacement
on page 57.
ERR
##
System Fault
Return for service. There are several error
codes. All error codes require return of
the device to an authorized service
center for repair. See
Return Procedure
on page 59.
RPL SEN
Defective sensor
Replace sensor.
The Masimo ReSposable
sensor system is
non-functional.
Replace the Masimo ReSposable sensor
system.