Masimo
Cleaning and Service Warnings and Cautions
WARNING
: Do not adjust, repair, open, disassemble, or modify Pronto. Injury to personnel or
equipment damage could occur. Return Pronto for servicing.
WARNING
: Electrical Shock Hazard: The battery should be installed and/or removed from the
Pronto by qualified personnel only.
WARNING
: Do not incinerate battery.
WARNING
: Do not use petroleum-based or acetone solutions, or other harsh solvents, to
clean the Pronto. These substances affect the device’s materials and device failure can result.
WARNING
: Do not touch, press, or rub the display panels with abrasive cleaning compounds,
instruments, brushes, rough-surface materials, or bring them into contact with anything that
could scratch the display.
CAUTION
: An operator may only perform maintenance procedures specifically described in
the manual. Refer servicing to qualified service personnel trained in the repair of this
equipment.
CAUTION
: Do not submerge Pronto in any cleaning solution or attempt to sterilize by
autoclave, irradiation, steam, gas, ethylene oxide or any other method. This will seriously
damage Pronto.
CAUTION
: Electric shock hazard: Do not open the Pronto cover except to replace the battery
or batteries.
CAUTION
: Electrical shock and flammability hazard: Before cleaning, always turn off the
device and remove batteries.
CAUTION
: Electrical Shock Hazard: Carry out periodic tests to verify that leakage currents of
patient-applied circuits and the system are within acceptable limits as specified by the
applicable safety standards. The summation of leakage currents must be checked and in
compliance with IEC 60601-1 and UL60601-1. The system leakage current must be checked
when connecting external equipment to the system. When an event such as a component
drop of approximately 1 meter or greater or a spillage of blood or other liquids occurs, retest
before further use. Injury to personnel could occur.
Note
: Excessive cleaning solution can flow into the monitor and cause damage to internal
components.