Pronto
Chapter 4: Operation
www.masimo.com
34
Masimo
Trend Data Format
After a successful download of the trend data, a .out file will be created containing the
trend-dump information in ASCII delimited format. The format is defined in the following
table:
Parameter
Specification
Date
MM\DD\YY
Time
HH:MM:SS
Installed Parameter/
Measurement
Numeric value (see the display ranges in the
Default Settings
on page
29)
Exception Messages
The exceptions are displayed as a 3-digit, ASCII encoded, hexadecimal
value. The binary bits of the hexadecimal value are encoded as follows:
000 = Normal operation; no exceptions
004 = Low Perfusion
400 = Low Signal IQ
800 = Masimo SET. This flag means the algorithm is running in full
Masimo SET® mode. It requires a Masimo SET® sensor and needs to
acquire some clean data for this flag to be set.
Sample Trend Output
11/30/11 00:09:36 SpO2=100 PR=070 PI=01.32 SpCO=00.00 Met=00.00 SpHb=15.1 PVI=000
EXC=00000000
12/01/11 00:12:45 SpO2=096 PR=068 PI=03.52 SpCO=00.00 Met=00.00 SpHb=13.8 PVI=000
EXC=00000000
12/01/11 00:13:14 SpO2=100 PR=069 PI=02.20 SpCO=00.00 Met=00.00 SpHb=14.3 PVI=000
EXC=00000000
12/02/11 00:09:27 SpO2=100 PR=068 PI=01.52 SpCO=00.00 Met=00.00 SpHb=14.2 PVI=000
EXC=00000000
12/02/11 00:10:58 SpO2=099 PR=071 PI=03.64 SpCO=00.00 Met=00.00 SpHb=15.8 PVI=000
EXC=00000000
12/02/11 00:15:04 SpO2=097 PR=068 PI=01.52 SpCO=00.00 Met=00.00 SpHb=10.8 PVI=000
EXC=00000000
12/03/11 00:11:31 SpO2=100 PR=072 PI=04.69 SpCO=00.00 Met=00.00 SpHb=12.5 PVI=000
EXC=00000000
Note
: Trend output data appears for the parameters/measurements noted above. Pronto only
stores output data for SpO2, PR, PI, and SpHb parameters/measurements.
SpCO, SpMet, and
PVI measurements are not available with Pronto
.
Note
: Pronto does not store continuous output data for SpO2, PR, PI, and SpHb. Pronto only
stores data related to SpHb, SpO2, PR, and PI during spot check tests results.