48
of
49|
P a g e
UserManual
6.2 Measurement Parameters and Accuracy
Parameter
Range
Resolution
Accuracy
mV
0-250.00 mV
0.01 mV
+0.02% of r 2 count
V
0-30.00 VDC
0.001 V
+0.02% of r 2 count
mA
0-24.00 mA
0.001 mA
+0.02% of r 2 count
6.3 Source Accuracy
Parameter
Range
Resolution Accuracy
mV
0-250.00 mV
0.01 mV
+0.02% of r 2 count
V
0-12.00 VDC
0.001 V
+0.02% of r 2 count
mA
0-24.000 mA
0.001 mA +0.02% of r 2 count
6.4 Testing Features
Feature
Value
Loop power output
24V DC, + 10% (24mA maximum)
Output impedance in HART
compatible mode
250Ω + 20%