UserManual
Parameter Name
Description / Options
Main Display
Select which Reading to be display as a Main Reading (Reading Displays in
Box in RUN Page).
Available Options:
Actual
Display the Actual Input Value
Percentage
Display the Percentage Value of the Input.
The Value depends on Input Range. These settings are
available from
MENU
MEASURE.
Scaled
Display the Scaled Value of the Input.
The Scale Value depends on Input Range, Input Scaled
Range & Transfer Function. These settings are available
from
MENU
MEASURE.
Filter(sec)
1
st
Order IIR Low Pass Filter for Input Reading.
Filter is useful when a measurement signal contains unwanted noise.
Range:
0.0 to 60.0 sec
Tare(unit)
The Tare value is subtracted from the reading of the measured value.
Here unit is changed according to current Input Type and Measure Display
Mode.
Range:
In accordance with Input Range & Measure Display Mode.
Note: Beware of the problems that may result in not seeing the true
measurement value
.
Additional Info.1
If Measure Mode is not selected as Display mode this parameter will be
disabled. And for Measure Display Mode, Reset Option is available.
Additional Info.2
NA
4.3.2Source Display Settings
This Page is appears in
RUN
MENU
DISPLAY
SOURCE.