User Manual
P a g e
|
45
of
49
Note:
Usually, LC-11 calibrates using the above steps in one/two time
only. But if the Input / Output are very out from the desired value
repeat the Calibration steps until the device calibrated properly.
4.6.5 Battery Info.
This page shows the basic battery Information.
Battery Level (Percentage)
Battery Voltage (in mV)
Current (in mA)
Battery Status
Time to Full (in min)
Time to Empty (in min)
4.6.6 Set Password
Change Device Password.
This password has to be entered before Calibrating & Resetting
Configuration Parameters.
To change password Enter Current Password. If the entered
password is correct then New Password EditBox will be enabled.
Then enter the New Password and press
MENU/ENT
key to store
it.
4.6.7Factory Reset
To Reset LC-11 Parameters to its Default Value.
To Reset Enter Current Password. If the entered password is
correct then New Password EditBox will be enabled. Then enter
the New Password and press
MENU/ENT
key to store it.
4.6.8 About Calibrator
This Page gives some basic information about the LC-11 hardware connection.