Alert Code/Name
Description/Action
Internal Balloon
Leak
Air Pump
Overactive
External Balloon
Not Deflating
143
144
145
The Console has detected a problem with balloon
inflation (internal).
• Ensure that the Cartridge Cable is not kinked or pinched
• Check for balloon inflation during pre-procedure test
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and is not leaking
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected a problem with the balloon inflation
system – the air reservoir pump is overactive.
• Ensure that the Cartridge Cable of the Water Vapor Probe
is not kinked or pinched
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and is not leaking
• If the issue persists, turn off the Console and contact an
AEGEA representative
The Console has detected a problem with external
balloon deflation.
• Ensure that the Cartridge Cable of the Water Vapor Probe
is not kinked, twisted or pinched
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
• Note: If the Water Vapor Probe is within the cervix,
disconnect the Cartridge from the Water Vapor Probe
Handle before removing the Water Vapor Probe from
the patient
• If the issue persists, turn off the Console and contact an
AEGEA representative
Alert Code/Name
Description/Action
Air Reservoir
Pressure Low
Air Reservoir
Pressure High
Balloon Test
Failure
146
147
148
The Console has detected a problem with the balloon inflation
system (low pressure issue).
• Ensure that the Cartridge Cable of the Water Vapor Probe
is not kinked or pinched
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and is not leaking
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
• If the issue persists, turn off the Console and contact an
AEGEA Medical representative
The Console has detected a problem with the balloon inflation
system (high pressure issue).
• Ensure that the Cartridge Cable of the Water Vapor Probe is
not kinked, twisted, or pinched
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe
• If the issue persists, turn off the Console and contact an
AEGEA Medical representative
The Console has detected that the Water Vapor Probe has failed
the balloon test.
• Check that the Protective Tip Cover is attached to the Water
Vapor Probe
• Check the Cartridge connection to the Water Vapor Probe
• Ensure that the Cartridge Cable of the Water Vapor Probe is
not kinked, twisted or pinched
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge