Alert Code/Name
Description/Action
Balloon Failure
in Thermal
Calibration
Middle Balloon
Not Deflating
Internal Balloon
Not Deflating
149
150
151
The Console has detected that the Water Vapor Probe has a
failed balloon.
• The Console may not allow the procedure to continue until
the Water Vapor Probe is replaced
• Check that the Protective Tip Cover is attached to the Water
Vapor Probe
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected a problem with middle balloon
deflation.
• Ensure that the Cartridge Cable of the Water Vapor Probe
is not kinked, twisted or pinched
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
• Note: If the Water Vapor Probe is within the cervix,
disconnect the Cartridge from the Water Vapor Probe
before removing the Water Vapor Probe from the patient
• If the issue persists, turn off the Console and contact an
AEGEA representative
The Console has detected a problem with internal balloon
deflation. The internal balloon is not deflating.
• Ensure that the Cartridge Cable of the Water Vapor Probe
is not kinked, twisted or pinched
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
• Note: If the Water Vapor Probe is within the cervix,
disconnect the Cartridge from the Water Vapor Probe
before removing the Water Vapor Probe from the patient
• If the issue persists, turn off the Console and contact an
AEGEA representative
Alert Code/Name
Description/Action
Middle Balloon
Inflation Too Slow
Internal Balloon
Inflation Too Slow
Pressure Sensor
Ambient Failure
152
153
154
The Console has detected a problem with the middle balloon
inflation. The middle balloon is taking too long to fill.
• Ensure that the Cartridge Cable is not kinked, twisted or
pinched
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and not leaking
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected a problem with the internal balloon
inflation. The internal balloon is taking too long to fill.
• Ensure that the Cartridge Cable is not kinked, twisted or
pinched
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected a problem with the intra-uterine
Pressure Sensor in the Water Vapor Probe. The Pressure
Sensor may be damaged.
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe
• Press “Clear” and allow the Console to re-evaluate the
Pressure Sensor system
• If the Console was stored in a hot or cold area, allow the
unit to stabilize to the room temperature environment
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge