13. Press green arrow on the Console
display to begin the Water Vapor
Probe Test and prepare it for use.
The Console screen will indicate
that the test is in progress and then
will indicate that the test passed
(images 8 and 9)
.
If the Water Vapor Probe test is
unsuccessful, the Console will indicate
an alert code.
⚠
Warning:
During the Water Vapor
Probe test, the three balloons on the
shaft of the Water Vapor Probe will
inflate. Visually confirm the inflation
of all three balloons. If one of the
balloons does not inflate, do not
proceed. Press the red Treatment
Interrupt Button to disconnect and
replace the Water Vapor Probe.
In addition to the balloon inflation
test, the Console will test the
intrauterine pressure sensor and the
saline flow system.
⚠
Warning:
During the Water Vapor
Probe test, the tip cover, distal
tip of the Vapor Probe and Water
Vapor Probe Outflow Tube may
become warm to the touch. Avoid
prolonged contact with personnel
or the patient.
11. Firmly insert the Console’s Cartridge
to the Water Vapor Probe handle
(image 6)
.
Tip
: Momentarily press the
Cartridge Release Button at the base
of the Water Vapor Probe handle to
facilitate insertion.
Note: Screen will automatically
advance when Cartridge is
connected properly.
12. Pull down on the Blue Tab at the base
of the Water Vapor Probe Handle
to pull the Sterile Sheath over the
Cartridge Cable
(image 7)
.