Table A1
⚠
Warning:
The emissions characteristics of this equipment make it
suitable for use in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B
is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user
might need to take mitigation measures, such as relocating or re-
orienting the equipment.
Audio Volume
40-75 dBa
Audible tone volume is adjustable on the first screen. However, audible
tones for Alarms are not adjustable.
⚠
Warning:
Auditory alarm signal sound pressure levels that are less than
ambient levels can impede Operator recognition of alarm Conditions.
Power Cord
Use AEGEA Medical Inc. provided power cords with the
Console
:
• Length, 8’
Technical Compliance Information
IEC 60601-1 Classifications:
• Type of protection against electric shock: Class I Equipment
• Degree of protection against electric shock: Type B Applied Parts
• Degree of protection against the ingress of water or particles:
– Console
IP21
– Cartridge IPX0
• Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
The Console needs special precautions regarding EMC (Electromagnetic
Compatibility) and needs to be installed and put into service according
to the EMC information provided in this Operator’s Manual. The Console is
intended to be used in a professional healthcare facility environment.
The testing results documented below are for the entire Vapor System
(GEA-SYS-16-0500), which includes the
Console
(GEN-16-050) and the
Water Vapor Probe (DDK-16-050).
Guidance and Manufacturer’s Declaration (Electromagnetic Emissions)
The GEA-SYS-16-0500 is intended for use in the electromagnetic environment specified
below. The customer or the user of the GEA-SYS-16-0500 should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic Environment –
Guidance
Conducted and
Radiated RF Emissions
CISPR 11
Harmonic Distortion
IEC 61000-3-2
Voltage Fluctuations and
Flicker IEC 61000-3-3
Group 1, Class A
Class A
Complies
The GEA-SYS-16-0500 uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
The GEA-SYS-16-0500 is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.