MARA OPERATOR’S MANUAL
MARA OPERATOR’S MANUAL
Table 2: Alert Codes 101-299
Alert Code/Name
Description/Action
Cervical
Temperature
Greater Than
Expected Body
Temperature
Water Vapor
Probe Cervical
Thermocouple
Fault
Syringe Attached
During Self-Test
Pressure Sensor
Not Zero
101
102
110
116
The Console has detected a high reading on the Cervical
thermocouple, indicating that the Water Vapor Probe may
be defective.
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
The Console has detected that the Water Vapor Probe is
defective (thermocouple issue) because the temperature
reading is too low.
• Depending upon the stage of the procedure, the Console
may not allow the procedure to continue until the Water
Vapor Probe is replaced
• Ensure that the Water Vapor Probe is not touching
other equipment
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions
to disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
The Console has detected that the Syringe is attached before
the appropriate step.
• Please remove the Syringe to allow the Self-Test to proceed
The Console has detected an issue with initializing the Water
Vapor Probe’s Pressure Sensor.
• Check that the Water Vapor Probe tip is not pressed
against a surface
• Check that the Protective Tip Cover is properly installed
on the Water Vapor Probe.
• Check that the Cartridge Cable is not impinged or pressed
against any objects
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
Alert Code/Name
Description/Action
Pressure Sensor
Calibration Error
Saline Prime
Error
Saline Re-prime
Error
Saline Prime
Pressure Error
117
118
119
120
The Console has detected that the Water Vapor Probe Pressure
Sensor calibration is not within expected range.
• Check that the Protective Tip Cover is properly installed
on the Water Vapor Probe
• Check the Cartridge connection to the Water Vapor Probe
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected that saline is not flowing as
expected.
• Check that the Vapor Probe tubing is connected to the
saline source
• Ensure that the tubing is not kinked or pinched
The Console has detected that saline is not flowing as
expected.
• Check that the Vapor Probe tubing is connected to the
saline source
• Ensure that the tubing is not kinked or pinched
The Console has detected a problem with the saline priming
process.
• Check that the Protective Tip Cover is properly installed
on the Water Vapor Probe
• Ensure that the tubing is not kinked or pinched
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge