MARA OPERATOR’S MANUAL
MARA OPERATOR’S MANUAL
Alert Code/Name
Description/Action
External Pressure
Low
IU Pressure High
After Treatment
External Pressure
Zero
Vapor Probe
Disconnected
Syringe Unloaded
208
211
219
229
230
The Console has detected that the intrauterine pressure is not
responding correctly and the pressure is too low.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear” and prepare to remove the
Water Vapor Probe from the patient
The Console has detected that the intrauterine pressure has
remained high after the treatment.
• Ensure that the conduits of the Water Vapor Probe are not
kinked or pinched, and specifically ensure the Water Vapor
Probe’s outflow line is not blocked or has its end
submerged in liquid
• Consider the potential for the uterus to be pressurized with
vapor due to a Console malfunction. Disconnect the
Cartridge to prevent the possibility of additional vapor
delivery.
• Press “Clear” to continue. Use caution when removing the
Water Vapor Probe from the patient to reduce any risk of
possible excess vapor from the distal end of the Water
Vapor Probe.
The Console has detected a very low intrauterine pressure. The
Water Vapor Probe positioning may have been compromised.
Water Vapor delivery has stopped and treatment has ended.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear”, wait 10 seconds, and then
remove the Water Vapor Probe from the patient.
The Cartridge has become detached from the Water Vapor
Probe.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear”, wait 10 seconds, and then
remove the Water Vapor Probe from the patient.
The Console has detected that the syringe was disconnected
and unexpectedly released from the Console.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear”, wait 10 seconds, and then
remove the Water Vapor Probe from the patient.
Alert Code/Name
Description/Action
Middle Balloon
Leak
Internal Balloon
Leak
Internal Balloon
Not Deflating
Treatment
Incomplete
(during Flush)
Treatment
Incomplete
(after Flush)
Syringe is Empty
RF Power Status
Wrong
242
243
251
261
262
264
271
The Console has detected a leak in the middle balloon.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds, and then
remove the Water Vapor Probe from the patient.
The Console has detected a leak in the internal balloon.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear”, wait 10 seconds, and then
remove the Water Vapor Probe from the patient.
The Console has detected a problem with internal balloon
deflation.
• If the Water Vapor Probe is within the cervix, disconnect
the Cartridge from the Water Vapor Probe before removing
the Water Vapor Probe from the patient
• If the issue persists, turn off the Console and contact an
AEGEA representative
The patient did not receive a full treatment. Water Vapor delivery
was automatically stopped during the flush. The procedure
may be re-started from the beginning after replacing the Water
Vapor Probe.
The patient did not receive a full treatment. Vapor delivery
was automatically stopped during water vapor treatment.
A second vapor treatment in the same operative setting must
not be attempted for this patient. A repeat ablation with the
Mara Water Vapor Ablation System has not been studied and
the effects are unknown.
The Console has detected that the syringe is empty.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear” and prepare to remove the Water
Vapor Probe from the patient.
The Console has detected that the RF Power has failed.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds and remove
the Water Vapor Probe from the patient.