MARA OPERATOR’S MANUAL
MARA OPERATOR’S MANUAL
Table 3 lists the steps for disconnecting or reconnecting the Water Vapor
Probe from the Cartridge or replacing the Water Vapor Probe if an issue
persists at various stages of the procedure.
Note:
If instructed to disconnect or reconnect the Water Vapor Probe from
the Cartridge or replace the Water Vapor Probe, be sure to wait 10 seconds
before taking the necessary action.
Alert Code/Name
Description/Action
Hypotube
Temperature
Error
Wire Temperature
Error
Cervical TC Error
Probe Temp Error
Pressure Sensor
Communication
Error
Interrupt While
Flushing
Interrupt While
Treating
280
281
284
289
290
298
299
The Console has detected that the Hypotube Temperature
Sensor has a Failure.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds and remove
the Water Vapor Probe from the patient.
The Console has detected that the Wire Temperature Sensor
has a Failure.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds and remove
the Water Vapor Probe from the patient.
The Console has detected a Cervical Temperature
Sensor Failure.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds and remove the
Water Vapor Probe from the patient.
The Console has detected multiple Temperature Sensor failures.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds and remove the
Water Vapor Probe from the patient.
The Console has detected that the intrauterine pressure
measurement function is not operational.
• Water vapor delivery has terminated early and treatment
has ended. Press “Clear,” wait 10 seconds and remove the
Water Vapor Probe from the patient.
• If the issue persists, replace the Console
The Console has detected that the Interrupt button has been
activated while Flushing.
• Twist the Interrupt button on the front of the Console to
Release the button
The Console has detected that the Interrupt button has been
activated during treatment.
• Twist the Interrupt button on the front of the Console to
Release the button
Table 3: Disconnecting and Reconnecting/Replacing
the Water Vapor Probe
Scenario
Action
Any Step Prior to Water
Vapor Probe Test
Any Step After Saline
is Delivered During
Water Vapor Probe
Test
Any Step After Water
Vapor Probe Insertion
Into the Uterus
1. Disconnect Water Vapor Probe from Cartridge
2. Clear error message(s)
3. Reconnect Water Vapor Probe to Cartridge
4. If issue persists, replace the Water Vapor Probe
1. Press Red Interrupt button, twist to release button
2. Choose “Quit” on screen
3. Disconnect Water Vapor Probe from Cartridge
4. Turn Console off
5. Turn Console on
6. Obtain new Water Vapor Probe
7. Follow on-screen instructions
1. Press Red Interrupt button, twist to release button
2. Choose “Quit” on screen
3. Remove Water Vapor Probe from the patient
4. Disconnect Water Vapor Probe from Cartridge
5. Turn Console off
6. Turn Console on
7. Obtain new Water Vapor Probe
8. Follow on-screen instructions