MARA OPERATOR’S MANUAL
MARA OPERATOR’S MANUAL
Alert Code/Name
Description/Action
EEPROM Write
Failure
Pressure Sensor
Optical
Connection
Failure
Pressure Sensor
Memory
Water Vapor
Probe
Disconnected
125
127
128
129
The Console has detected a memory failure in the Water
Vapor Probe.
• The procedure can continue; there is no need to replace
the Vapor Probe. Procedural data may not be recorded in
the Vapor Probe’s memory
The Console has detected that the Water Vapor Probe is
defective or not fully connected.
• Check that the Cartridge is firmly inserted
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
The Console has detected that the Water Vapor Probe is
defective or not fully connected.
• Check that the Cartridge is firmly inserted
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
The Console has detected the disconnection of the Water
Vapor Probe pressure sensor.
• Depending upon the stage of the procedure, the Console
may not allow the procedure to continue until the Water
Vapor Probe is replaced
• Check that the Cartridge is firmly attached to the Water
Vapor Probe
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
Alert Code/Name
Description/Action
Syringe Unloaded
External Balloon
Inflation Too Slow
External Balloon
Leak
Middle Balloon
Leak
130
140
141
142
The Console has detected that the Syringe was unexpectedly
released/disconnected.
• Reinstall the syringe to resume treatment workflow
• Depending upon the stage of the procedure, the Console
may not allow the procedure to continue until the Water
Vapor Probe is replaced
The Console has detected a problem with balloon inflation
(external). The balloon is taking too long to fill.
• Ensure that the Cartridge Cable is not kinked or pinched
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and is not leaking
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected a problem with balloon inflation
(external).
• Ensure that the Cartridge Cable is not kinked or pinched
• Check for balloon inflation during pre-procedure test
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and is not leaking
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
The Console has detected a problem with balloon
inflation (middle).
• Ensure that the Cartridge Cable is not kinked or pinched
• Check for balloon inflation during pre-procedure test
• Ensure that the Cartridge is firmly connected to the Water
Vapor Probe and is not leaking
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge