MARA OPERATOR’S MANUAL
MARA OPERATOR’S MANUAL
Alert Code/Name
Description/Action
Probe Lifetime
Exhausted
Probe Temp Error
Pressure Sensor
Communication
Alert
Water Vapor
Probe Attached
During Self-test
Water Vapor
Probe
Disconnected
188
189
190
191
197
The Console has detected that the Water Vapor Probe is being
re-used. Reuse of the Water Vapor Probe is not allowed.
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected multiple Temperature Sensor failures.
The Water Vapor Probe is defective or not fully connected.
• Clear error message(s).
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected that the intrauterine pressure sensor
is not being measured correctly.
• Upon clearing the alert, the Console will try to restore
functionality
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor Probe
from the Cartridge
The Console has detected that the Water Vapor Probe is
attached to the Cartridge before the appropriate step.
• Disconnect the Water Vapor Probe from the Cartridge to
allow the Self-Test to proceed
The Console has detected the disconnection of the Water
Vapor Probe.
• Clear error message(s)
• If the issue persists, refer to Table 3 for instructions to
disconnect and reconnect/replace the Water Vapor
Probe from the Cartridge
• If the issue persists, turn off the Console and contact an
AEGEA Medical representative
Alert Code/Name
Description/Action
Interrupt Button
Engaged
Cervical
Temperature
Greater Than
Expected Body
Temperature
Cervical
Thermocouple
Circuit Fault
External Pressure
Does Not Rise
External Pressure
High
External Pressure
Drop
199
201
202
205
206
207
The Console has detected that the Interrupt Button is Engaged.
• Release the Interrupt button by rotating it clockwise to
release the Interrupt button and proceed
- Press Quit to end the procedure and/or replace
the Water Vapor Probe
- Press Clear to resume
The Console has detected a cervical temperature higher
than expected.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear” and prepare to remove the
Water Vapor Probe from the patient
The Console has detected that the Water Vapor Probe
temperature sensor is defective.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear” and prepare to remove
the Water Vapor Probe from the patient
The Console has detected that the intrauterine pressure is not
responding correctly or the pressure is not rising.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear” and prepare to remove the
Water Vapor Probe from the patient
The Console has detected that the intrauterine pressure
is too high.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear” and prepare to remove the
Water Vapor Probe from the patient
The Console has detected that the intrauterine pressure is not
responding correctly and/or a sudden pressure drop could
indicate a leak in the uterus or Water Vapor Probe.
• Water vapor delivery has stopped and treatment has
terminated early. Press “Clear” and prepare to remove
the Water Vapor Probe from the patient