RS-232C INTERFACE - Section 7
7-16 MTS-3000 SERIES OPERATION AND REFERENCE MANUAL
#
0$17$
#
7(67
#
6<67(06
############################################
CU M004 01A
7.2.12 Digital Input/Output and DC Current Control
This section only applies to units with the DIO option.
The following commands are associated with control of the DIO module:
Command
Description
DIO x y z
Set the specified DIO settings
•
Where x = DIO module number: x = 0 for first DIO module, and 1 if referring to
the second DIO module if fitted
•
Where y = DIO setting code:
•
y = 101 for debounce time
•
y = 102 for DC current setting
•
y = 103 for DC current enable
•
y = 104 for Input Channel 1 trigger mode
•
y = 106 for Input Channel 2 trigger mode
•
y = 108 for tone control
•
y = 109 for output 1 contact delay
•
y = 110 for output 1 contact mode
•
Where z = DIO setting value:
•
z = debounce time in ms, valid values are 0-999 (y = 101)
•
z = DC current in amps, valid values are 0-2.5 (y = 102)
•
z = 1 for DC enabled, 2 for DC disabled (y = 103)
•
z = 1 for input trigger type OFF, (y = 104 or 106)
•
z = 2 for input trigger type FAULT, (y = 104 or 106)
•
z = 4 for input trigger type PULSE TIMING, (y = 104 or 106)
•
z = 8 for input trigger type TMR START, (y = 104 or 106)
•
z = 16 for input trigger type TRIP, (y = 104 or 106)
•
z = 32 for input trigger type FOOTSWITCH, (y = 104 or 106)
•
z = 64 for input trigger type RESET, (y = 104 or 106)
•
z = 128 for input trigger type RECLOSE, (y = 104 or 106)
•
z = 1 for TONE enabled, 0 for TONE disabled (y = 108)
•
z = output contact delay time in ms, valid values are 0-9999 (y = 109)
•
z = 1 for output contact type FAULT, (y = 110)
•
z = 2 for output contact type BREAKER-OPEN, (y = 110)
•
z = 4 for output contact type PERMISSIVE, (y = 110)
•
z = 8 for output contact type UNBLOCK, (y = 110)
•
z = 16 for output contact type OFF, (y = 110)
•
z = 32 for output contact type FAULT-NC, (y = 110)
•
z = 64 for output contact type RESET, (y = 110)