RS-232C INTERFACE - Section 7
MTS-3000 SERIES OPERATION AND REFERENCE MANUAL 7-15
CU M004 01A
###########
0$17$
#
7(67
#
6<67(06
•
Kernal V0.4.03 > Kernal version
•
Mar 1 2001 > Firmware release date
•
1 = MTS-3000
•
2 = MTS-3010
•
3 = MTS-3020
•
4 = MTS-3030
•
5 = MTS-3040
•
6 = MTS-3060
FIA# Fault Incidence Angle
•
Set fault incidence angle control
•
Valid values are 0 - 360 (degrees)
•
No value sets FIA control to random (off)
DEF Default settings
•
Restores all settings to power-up default values
•
See Section 5.2.1 for details
! Repeat Last Command
•
One character repeat command to save transmission time
CWN Clear warnings
•
This command clears all latched alarm status bits. This command is typically
executed prior to a test, after the test has completed the PWS command is
executed to make sure no alarm bits were latched during the test.
7.2.11 DC Voltage Control
This section only applies to units with the DC Voltage output option.
The following commands are associated with control of the DC voltage output:
Command Description
DCV x y z Set the specified DC voltage parameter to the specified value
•
Where x = 0
•
Where y = command:
•
y = 122 for set prefault voltage, z = 0 - 300
•
y = 123 for set fault voltage, z = 0 - 300
•
y = 124 for set postfault voltage, z = 0 - 300
•
y = 120 for output enable / disable, z = 0 to disable, z = 1 to enable
VDC# DC Voltage
•
Sets the DC voltage amplitude for the present fault state
•
Valid values are 0 - 300. (volts)