APPENDIX C
SUMMARY OF ERROR MESSAGES
The following is a summary of the most common error or alarm messages occurring with the MTS-3000.
Each of these messages is accompanied by a distinctive long error "Beep". Alarm messages are displayed
on the module where the fault condition is detected. Pressing any key will clear the error/alarm message.
If the error condition is still present, or re-occurs, the message and long "Beep" will likewise re-occur.
Note that, when you press a key to clear an error/alarm, the keypress also will cause whatever action that
keypress normally would cause. If you want to clear the error/alarm without causing any other action, a
good key to press is the "CE" key, since this key normally doesn't do anything except beep.
As noted in (C) below, any error/alarm message not listed here is usually an indication of design-level
problems with the MTS-3000, and the manufacturer should be notified, along with the particulars.
C.1 USER ERRORS - INAPPROPRIATE BUTTON PRESSING
Chg Flt Type 1st: The user tried to perform some action while in a fault type that’s inappropriate for
that particular action. To avoid this alarm, first change to a suitable fault type by
entering the following key sequence: MENU | Settings | Fault Type |.
Chg to Vectr 1st: The user tried to perform some action requiring Vector fault type while the system
was in some other fault type. To avoid this alarm, first change to the Vector fault
type by entering the following key sequence: MENU | Settings | Fault Type | Vector.
Chg VI Cnfig 1st: The user tried to perform some action that’s not compatable with the present VI
module amplifier configuration (e.g selecting
Φ
-
Φ
fault type when there’s only one
ACV channel and one ACI channel available). To avoid this alarm, first change to a
suitable configuration by entering the following key sequence: MENU | Preferences
| VI Module #n | Ch n.
Exceeded Limits: The keyboard data just entered is outside the acceptable range for the quantity being
set. Press CE and try again.
Finish # first: The user tried to perform some action while in the middle of entering data for
setting some quantity. The data entry must be completed (or terminated by pressing
CE, CE) before this action can be undertaken.
No Qty Selected: The user tried to enter a setting without first selecting a quantity for adjustment.
Select a quantity and try again.
Press ENABLE 1st: The user tried to perform some action while the AC outputs were inappropriately
enabled or disabled. To avoid this alarm, first change the status of the AC outputs
(by pressing ENABLE), and then try again.