System settings to activate AutoCal
6
Waiting period until the measured value of the span gas is displayed on the screen
(HSS3).
7
Waiting period, RS1 AutoCal, until analyzer is in operating mode again.
5
6
4
7
Fig. 80: AutoCal activation: change system settings (second part of the list)
4
Option AutoCal test mode
5
Waiting period HSS2 zero gas
6
Waiting period HSS3 span gas
7
Waiting period RS1 AutoCal
After switching from manual calibration to AutoCal, the display changes to: ‘Please re-start
the device OFF and ON again!’.
Note
The analyzer needs to be restarted after changing the parameters.
Start AutoCal procedure
Please tap on the calibration button. The M5/S1 screen opens. Scroll to ‘AutoCal’ on the
scroll bar at the left hand side.